Chapter 11 Page no. 1804
JC-PM-A-425.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A425 - IDC Air Filter Cleaning
6 PROCEDURE
To clean the IDC air filter, proceed as follows:
1. Remove the screws (1) securing the air filter cover (2).
2. Remove the air filter, and use a vacuum cleaner to ensure that the air filter is free from dust.
3. Put back the air filter into the air filter cover (2).
4. Use the two screws to secure the both the air filter cover and air filter to the IDC framework.