JC-LOTO-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card LOTO A001 - Lock-Out / Tag-Out (LOTO) Procedures
8-4 Isolation of the Generator Line Supply power source
You must isolate two types of power source which lead into the Generator of the Senographe system:
• the Mains Distribution Panel which supplies the power source for the Line Supply Cable going into
the Generator
• the power source for the room lamps running into the Generator
8-4-1 Isolating the Mains Distribution Panel
The Mains Distribution Panel may vary in different hospital, and the specific details are listed in the site
specific LOTO procedure. Use the following generic steps to isolate the power source going to the
Senographe system:
1. Switch off the AC mains power going to Generator cabinet. To do this, turn the handle of the Mains
Distribution Panel into the OFF position.
2. Press the OFF button on the front of the Mains Distribution Panel where fitted.
3. Turn all necessary switches/breakers on the front of the Mains Distribution Panel into the OFF posi-
tion.
4. If possible, disconnect the AC mains power cable from the Mains Distribution Panel.
5. If there exists a connection between the socket on the Mains Distribution Panel and the entrance into

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals