Job Card LOTO A001 - Lock-Out / Tag-Out (LOTO) Procedures
Page no. 387
JC-LOTO-A-001.fm
9 PROCEDURE FOR THE RECONNECTION OF THE EQUIPMENT
Follow this procedure to return the Senographe system or subsystem components to its normal opera-
tion
1. Inform everybody of the reconnection of the device.
2. Check the area and the equipment.
3. Make sure that all the parts and covers are correctly fitted.
4. Remove the tools from the area of the system.
5. Remove the LOTO devices including locking pins or bolts.
6. Re-establish the power to the system.
7. Return the equipment in its normal operating state.
8. Perform a system function check.

Table of Contents