Chapter 6 Page no. 398
JC-PHY-A-002.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A002 - Unpacking the Control Station
6-2 Removal from Pallet
1. Unpack the equipment.
2. Install the long wooden ramp (1), coming from column packaging, to the right hand side of the Con-
trol Station (the side that is connected to the Gantry).
3. On the left hand side of the Control Station, remove the three hex bolts (2) that secure the bracket to
the Control Station base.
1
2