EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 Page no. 582
JC-ELE-A-015.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A015 - Gantry Functional Checks
compression force between 5 kg and 15 kg.
4. Repeat the check as described above, but using a 50 mm thick acrylic plate.
The displayed value must now remain in the range 50 mm ±2 mm for any value of compression force
between 5 kg and 15 kg.
5. If either or both of the above checks gives an unsatisfactory result, perform the procedures described
in Job Card CAL A050 - Compression Thickness Calibration on page 1721.
6-10 Check Stop Motion Buttons
Check the red stop motion buttons (located on either side of the Gantry column base).
Both of these buttons must be tested:
- Press one of the stop motion buttons and check that all movement of the Gantry is stopped.
- Restart the Gantry. Press the other stop motion button and check that all movement of the Gantry
is stopped.

Table of Contents

Other manuals for GE Senographe DS

Related product manuals