Chapter 7 Page no. 582
JC-ELE-A-015.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A015 - Gantry Functional Checks
compression force between 5 kg and 15 kg.
4. Repeat the check as described above, but using a 50 mm thick acrylic plate.
The displayed value must now remain in the range 50 mm ±2 mm for any value of compression force
between 5 kg and 15 kg.
5. If either or both of the above checks gives an unsatisfactory result, perform the procedures described
in Job Card CAL A050 - Compression Thickness Calibration on page 1721.
6-10 Check Stop Motion Buttons
Check the red stop motion buttons (located on either side of the Gantry column base).
• Both of these buttons must be tested:
- Press one of the stop motion buttons and check that all movement of the Gantry is stopped.
- Restart the Gantry. Press the other stop motion button and check that all movement of the Gantry
is stopped.