Chapter 7 Page no. 590
JC-ELE-A-020.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7 PRINTER PARAMETERS
This section gives information on entries to be made in the DICOM Parameters window for printers qual-
ified for use with the Senographe system.
7-1 Konica Printers
7-1-1 Konica 793 DICOM Printer
• DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g., Konica
DPRO793.
• Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., dpro793.
• Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.48.16.135.
• Application Entity Title (AET): Must be DPRO793
• Port number: Must be 104
• Layouts: Must be 1 x 1 , 1 x 2 , 2 x 1 , or 2 x 2
• Film size: Must be 8 in x 10 in W = 4596 H = 5674
10 in x 12 in W = 5766 H = 6826
• Pixel size 43.75
• Configuration information Must be KC_ LUT=1
• Density Must be Min: 20 Max: 400
• Magnification type: Must be Cubic
• Smoothing factor: Must not contain any value
• Trim: Must be NO
• Printer Memory Size Must be 1024

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals