Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-1-2 Konica 873 DICOM Printer
DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g., KM-
DRYPRO873.
Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., dpro873.
Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.48.16.135.
Application Entity Title (AET): Must be dpro873
Port number: Must be 104
Layouts: Must be 1 x 1 , 1 x 2 , 2 x 1 , or 2 x 2
Film size: Must be 8 in x 10 in W = 4596 H = 5674
10 in x 12 in W = 5766 H = 6826
Pixel size 43.75
Configuration information Must not contain any value
Density Must be Min: 20 Max: 400
Magnification type: Must be Cubic
Smoothing factor: Must not contain any value
Trim: Must be NO
Printer Memory Size Must be 1024