Chapter 7 Page no. 596
JC-ELE-A-020.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-2-2 Agfa 5500/5503 DICOM Printer
• DICOM printer label: Any name you wish to see appearing on the Film Composer e.g., Drystar5503.
• Hostname: Use printer Hostname if any, or a name unique to your Network e.g., drystar5503.
• Network address: A unique IP address given by the hospital Network Administrator, e.g.,
3.48.16.136.
• Application Entity Title (AET):Default value is DS5500 , however contact the Agfa engineer for
the correct AET value to use
• Port number: Must be 104
• Layouts: Must be 1 x 1 , 1 x 2 , 2 x 1 , or 2 x 2
• Film size: Must be 8 in x 10 in W = 3828 H = 4958
10 in x 12 in W = 4892 H = 5900
• Pixel size 50
• Configuration information:Must be PERCEPTION_LUT=OEM919
• Density Must be Min: no value Max: 370
• Magnification type: Must be Cubic
• Smoothing factor: Must be 143
• Trim: Must be NO
• Printer Memory Size Must be 100