Chapter 7 Page no. 600
JC-ELE-A-020.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-3 Kodak / Carestream Printers
7-3-1 DryView 8900 DICOM Printer
DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g., Kodak 8900.
Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., DV8900.
Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.48.16.135.
Application Entity Title (AET): Must be 8900
Port number: Must be 5040
Layouts: Must be 1 x 1 , 1 x 2 , 2 x 1 , or 2 x 2
Film size: Must be 8 in x 10 in W = 5056 H = 6368
10 in x 12 in W = 6368 H = 7656
Pixel size 39.08
Configuration information Must be LUT=gsdfaa0p.w8u,3
Density Must be Min: no value Max: 360 for DVM film type
Max: 415 for DVM+ film type
Magnification type: Must be Cubic
Smoothing factor: Must be 15
Trim: Must be NO
Printer Memory Size 40 (default value)
Note:
To match new AFSSAPS requirements (in France) for Ambient Light=1 & Illumination=4000, do
not set Review Room into Review Room Management, and check with the printer manufacturer
that the printer is correctly configured.