Chapter 7 Page no. 616
JC-ELE-A-020.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-5 Sony Printers
7-5-1 SONY UP-DF750 Printer
DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g., SONY UP-
DF750.
Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., DF750.
Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.48.16.135.
Application Entity Title (AET): Must be UPDF750M
Port number: Must be 104
Layouts: Must be 1 x 1 , 1 x 2 , 2 x 1 , or 2 x 2
Film size: Must be 8 in x 10 in W = 4736 H = 5856
10 in x 12 in W = 5952 H = 7040
Pixel size 42.05
Configuration information Must not contain any value
Density Must be Min: 20 Max: 380
Magnification type: Must be Cubic
Smoothing factor: Must not contain any value
Trim: Must be NO
Printer Memory Size 20
Note:
In order for the Sony DP-DF750 to work with the GE mammography image size produced by the
Senographe system, the printer default settings must be changed during the installation. The de-
fault Frame Type value of “Caption2” must be changed to “M-Wide2” by the Sony printer techni-
cian.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals