Chapter 7 Page no. 632
JC-ELE-A-021.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
region. Press <Return> to confirm the selection, and then press the <F2> key to continue.
13. At the Offset from GMT screen that appears, type 0 for the offset the press <Return> to confirm the
offset value. Press the <F2> key to continue.
Time
Zone
On this screen .................................
> To make a selection .............................
Regions :
X
Africa
Asia, Eastern
United States
--
-
Other - Offset from GMT
Other

Specify the number of hours difference between ....
0


Offset From GMT