EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #634 background imageLoading...
Page #634 background image
Chapter 7 Page no. 634
JC-ELE-A-021.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
7-3 Network Configuration for V3 Control Stations
For V3 Control Stations, proceed as follows to complete the network configuration.
1. At the Network Connectivity screen, ensure that Yes is selected for the Networked option, and press
the <F2> key to continue.
2. At the Configure Multiple Network Interface screen, ensure that only nge0 is selected from the list of
Network Interfaces. Press the <Return> key to confirm the selection, and then press the <F2> key to
continue.
3. At the DHCP for nge0 screen ensure, that No is selected for the Use DHCP for nge0 option, and
press the <F2> key to continue.

Specify Yes if the system is connected to ..................
Networked
Yes
No
X
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Network Connectivity
......................................................
X

Multiple network interfaces have been ..................
Network Interfaces
nge0
nge1
‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐
Note : You must choose at least one interface to configure.
Configure Multiple Network Interfaces
X

Specify whether or not ...
Use DHCP for nge0
Yes
No
X

DHCP for nge0
‐‐‐‐‐‐‐‐‐‐
X

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals