Chapter 7 Page no. 676
JC-ELE-A-024.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A024 - Declare Performed Procedure Step Hosts
6 PROCEDURE
1. From the Browser, click the Tools menu button , then click PPS management.
The PPS Management selection window like the following appears.
• This window is divided into two sections:
- The upper section is used to declare the PPS remote servers associated with the archiving sys-
tem and/or the CAD system.
- The lower section is used for management of the PPS retry mechanism if the remote server
returns an error.
Note:
The PPS Server Application Entity Title, IP address, and port number are attached to a specified
Image Archiver or CAD system. You must therefore select a system from the drop-down menu be-
fore entering PPS configuration data; otherwise the changes are lost.
• Determine the number of remote PPS servers (one or two) you have to declare, according to the site
workflow. Each PPS instance references the images that have been autopushed to the associated

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals