Chapter 7 Page no. 704
JC-ELE-A-032.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A032 – IQ Tools Flat Field Test
9-5 Final checks
!
Notice:
If a new gain calibration was done, you are recommended to recalibrate all other gain configura-
tions.
After investigating and correcting the suggested causes of error, repeat the test.
If the test still fails, call your OLC.