GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
Page no. 849 Chapter 9
ERR-SUB-A-001.fm
096/024 I06 Abort: Detect
sig. out range
Dose value returned by IDC out of
range
Reboot the full system.
096/025 I08 Track/filter
change
Track/filter configuration changed
between the pre-shot and the exposure
None (warning only).
096/033 I11 IDC sent an invalid dose value to the
Senographe.
Reset the entire system and retry the
exposure.
096/038 I31 Abort: Image
signal too low
Image signal too low. Occurs in AOP mode, between pre-
exp and exposure, if the dose sent by
IDC is out of permitted range.
Check:
Mechanical thickness.
Paddle type used.
Presence of implant or marking in
the breast.
That paddle is located in the center
of the trolley support.
That the paddle type is not changing
during compression (paddle label
position not correct).
096/039 I32 Abort: Image
signal too high
Image signal too high. Occurs in AOP mode, between pre-
exp and exposure, if the dose sent by
IDC is out of permitted range.
Check:
Mechanical thickness.
Paddle type used.
Presence of implant or marking in
the breast.
That paddle is located in the center
of the trolley support.
That the paddle type does not
change during compression (paddle
label position not correct).
096/040 I33 Abort No AOP
zone
Exposure aborted; no AOP zone estab-
lished.
Check for presence of breast implant.
096/041 I38 Thermal limit:
mAs not optimal
Due to the system thermal condition,
the exposure mAs value has been
reduced to be within the regular Image
Quality limits
.
None (warning)
096/042 I39 Exp. time limit:
mAs not optimal
Due to exposure time limit, the expo-
sure mAs value has been reduced to be
within the regular Image Quality limits.
None (warning)
096/043 I40 Exp. time limit:
risk mAs too low
Maximum exposure time reached; the
exposure mAs value has been reduced
more than the regular Image Quality
limits.
None (warning)
Code Message text Description Recommended actions

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals