ERR SUB A003 - Gantry Error Messages
161/012 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The lift motion was stopped because the Gantry
CPU received an inconsistent UIF request.
Operator error. No service recommended actions.
161/013 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The rotation motion was stopped because the Gan-
try CPU received an inconsistent UIF request.
Operator error. No service recommended actions.
161/014 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The rotation UIF request was rejected by the Gantry
CPU because a compression motion is currently
active.
Operator error. No service recommended actions.
161/015 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The rotation UIF request was rejected by the Gantry
CPU because a lift motion is currently active.
Operator error. No service recommended actions.
161/016 ADVISORY LCD: A02 Compres-
sion force is applied
The rotation UIF request was rejected by the Gantry
CPU because a compression force is currently
applied.
Operator error. No service recommended actions.
161/017 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The lift UIF request was rejected by the Gantry CPU
because a compression motion is currently active.
Operator error. No service recommended actions.
161/018 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The lift UIF request was rejected by the Gantry CPU
because a lift motion is currently active.
Operator error. No service recommended actions.
161/019 ADVISORY LCD: A02 Compres-
sion force is applied
The lift UIF request was rejected by the Gantry CPU
because a compression force is currently applied.
Operator error. No service recommended actions.
161/020 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The compression UIF request was rejected by the
Gantry CPU because a rotation motion is currently
active.
Operator error. No service recommended actions.
161/021 ADVISORY LCD: A10 Motion
stopped. Multiple
movements are
requested
The compression UIF request was rejected by the
Gantry CPU because a rotation lift is currently active.
Operator error. No service recommended actions.
161/022 ADVISORY LCD: A15 Rotation
end position is
reached
The rotation UIF request was rejected by the Gantry
CPU because it is already at its end position.
Operator error. No service recommended actions.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card