GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 929 Chapter 9
ERR-SUB-A-003.fm
3-8 Error codes from 170/yyy: Bucky
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card
170/000 INHIBIT GRID NODE
ERROR
This error is linked to an invalid command sent by
Poseidon (exposure time out of range for instance).
1. Reboot the system from the X-ray Console.
2. If the problem persists, check the CAN bus
Poseidon-Bucky.
170/001 DEBUG
ONLY
GRID NODE
ERROR
Command not processed because the previous com-
mand is being processed.
Ignore this error.
170/002 INHIBIT GRID NODE
ERROR
This error is linked to a system CAN overrun, object
lost.
1. Reboot the system from the X-ray Console.
2. If the problem persists, check the overall system
CAN bus.
170/003 INHIBIT GRID NODE
ERROR
This error is due to an invalid PDO length.
1. Reboot the system from the X-ray Console.
2. If the problem persists, check the overall system
CAN bus.
3. If the problem persists, change the Bucky and
check the mechanical thickness.
D/R A255
page
1339
ELE A015
page 579
170/004 INHIBIT GRID NODE
ERROR
The Gantry CPU heartbeat was not received.
1. Reboot the system from the X-ray Console.
2. If the problem persists, check whether this failure
is reported on other nodes meaning that Gantry
CPU does not send system heartbeats. Hence
replace Gantry CPU board (re-calibrate the colli-
mator afterwards).
3. If the error 162/007 is raised, check Poseidon-
Stereotix CAN bus.
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
170/005 DEBUG
ONLY
GRID NODE
ERROR
This error is due to a correlation error and is linked to
a Bucky step lost error.
1. Reboot the system from the X-ray Console.
2. If the problem persists, change the Bucky and
check the mechanical thickness.
Note: If this error occurs more than five times per
month, change the Bucky.
D/R A255
page
1339
ELE A015
page 579