EasyManua.ls Logo

Hioki RM3545 - 8.3 Multiplexer Settings

Hioki RM3545
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.1 About the Multiplexer
140
Two Z3003 Multiplexer Units can be installed on the RM3545-02.
Number of locations that can be measured
Benefits of using the Multiplexer Unit
Since each channel’s pins can be combined freely, it is possible to measure a variety of
measurement targets with the minimum amount of wiring.
See: "8.7 Example Connections and Settings"(p.167)
Example: 3-phase motor with Δ wiring or Y wiring
Series elements such as a network resistor
Independent elements
Different measurement conditions can be set for each channel.
See: "8.3 Multiplexer Settings"(p.148)
Batch zero-adjustment can be performed for the desired channels.
See: "8.5 Zero Adjustment (When a Multiplexer Unit Has Been Installed)"(p.163)
Judgments can be made using measured values as references.
See: "Setting Basic Measurement Conditions and Total Judgment Conditions for Individual Chan-
nels" (p.156)
Up to 42 channels can be registered.
Up to eight setting panels (panel number: 31 to 38) can be saved, apart from measure-
ment conditions for which the multiplexer is not used (when using the measurement ter-
minals on the front of the instrument).
You can choose from the following three scan methods according to your application:
(1) Scan function: OFF
(2) Scan function: Step
(3) Scan function: Auto
8.1 About the Multiplexer
Number of units 2-wire 4-wire
1 unit 21 locations 10 locations
2 units 42 locations 20 locations