4.5 Correcting for the Effects of Temperature (Temperature Correction (TC))
75
4
Temperature correction converts resistance values to resistance values at standard tem-
perature and displays the result.
For more information about the principle of temperature correction, see "Appendix 4 Tem-
perature Correction (TC) Function" (p. A4).
To perform temperature correction, connect the temperature sensor or thermometer with
analog output to the TEMP. jack on the rear of the instrument.
See: "2.3 Connecting Z2001 Temperature Sensor or Thermometer with Analog Output (When
using the TC or T)" (p.37)
See: "3.1 Checking the Measurement Target" (p.48)
Setting T to on causes TC to be turned off automatically.
4.5 Correcting for the Effects of Temperature
(Temperature Correction (TC))
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Enable the temperature correction function. (TC)
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
2
Enables the TC function
Disables the TC function
(default) (go to step 5)
Selection
1