4.11 Improving Probe Contact (Contact Improver Function)
90
Probe contacts can be improved by applying current from the SENSE A to the SENSE B
probes before measuring.
The maximum contact improvement current is 10 mA, and the maximum applied voltage is 5 V.
When low power is set to on, the contact improver function is set to off.
Using the contact improver function causes the time until measurement completion to be
lengthened by 0.2 ms.
Timing Chart (Contact Improver Function)
4.11 Improving Probe Contact (Contact Improver
Function)
The Contact Improver function applies voltage to the sample. Be careful
when measuring samples with characteristics (magnetoresistive elements,
signal relays, EMI filters, etc.) that may be affected.
1
Open the Settings Screen.
Probe contact condition
Contact Improver current
Measurement current
(10 m to 1,000 range)
Measurement current
(10 k to 1,000 M range)
Measurement
Delay
Measuring
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2