5.3 Performing Statistical Calculations on Measured Values
111
5
Statistical calculations can be performed on up to 30,000 measured values, with results dis-
played. Printing is also available (p. 245).
Calculation types: average, maximum and minimum values, population standard deviation,
sample standard deviation, process compatibility indices
Deleting Statistical Calculation Results
Stored data is automatically erased at the following times:
5.3 Performing Statistical Calculations on
Measured Values
Maximum value
Xmax = MAX (x
1
, ....., x
n
)
In these formulas, n represents the number
of valid data samples.
* Process capability index
The process capability indices repre-
sent the quality achievement capability
created by a process, which is the
breadth of the dispersion and bias of
the process' quality. Generally,
depending on the values of Cp and
Cpk, process capability is evaluated as
follows:
Cp, Cpk >1.33............Process capabil-
ity is ideal
1.33 Cp, Cpk >1.00.Process capabil-
ity is adequate
1.00 Cp, Cpk ...........Process capabil-
ity is inadequate
UPP and LOW are the upper and
lower thresholds of the comparator.
When the BIN function is on, the pro-
cess capability index will not be calcu-
lated.
Minimum value
Xmin = MIN (x
1
, ....., x
n
)
Average
Population
standard
deviation
Standard
deviation
of sample
Process
capability index
(dispersion) *
Process
capability
index (bias)*
n
x
x
=
σ
n
x
2
nx
2
n
---------------------------=
σ
n 1
x
2
nx
2
n 1
---------------------------
=
Cp =
UPPLOW
6σ
n1
Cpk =
UPPLOW UPP + LOW2x
6σ
n1
When only one valid data sample exists, standard deviation of sample and process capa-
bility indices are not displayed.
When σ
n-1
= 0, Cp and Cpk are 99.99.
The upper limit of Cp and Cpk is 99.99. If Cp or Cpk exceeds 99.99, the value 99.99 is dis-
played
Negative values of Cpk are handled as Cpk = 0.
If statistical calculation is turned off and then back on without first clearing calculation
results, calculation resumes from the point when it was turned off.
Measurement speed is restricted when statistical calculation is enabled.
Turning on T or setting the measurement terminal to multiplexer automatically turns off
the statistical calculation function.
when changing measurement conditions (low-power, measurement current, OVC,
100 M range high-precision mode, TC, non-offset scaling settings)
when changing comparator settings (p.98)
when changing BIN measurement function settings (p.108)
when printing the statistical calculations (p. 245) (you can select whether to delete
results after printing (p.246))
upon system reset (p. 134)
when turning off the instrument

Table of Contents

Other manuals for Hioki RM3545

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki RM3545 and is the answer not in the manual?

Hioki RM3545 Specifications

General IconGeneral
BrandHioki
ModelRM3545
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals