EasyManua.ls Logo

Hioki RM3545 - Measurement

Hioki RM3545
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.5 Checking Measured Values
57
3
*1 Over-range Detection Function
Examples of Over-range Faults
*2 Current Fault Detection Function
Example of Current Fault
SOURCE A or SOURCE B probe open
Broken measurement target (open work)
SOURCE A or SOURCE B cable break, poor connection
SOURCE wiring resistance in excess of the following values may cause a current fault,
making measurement impossible. When using measurement current 1 A ranges, keep the
wiring resistance as well as the contact resistance between the measurement target and
measurement lead low.
Over-range Detection Measurement Example
The measured value is outside of
the measurement range.
Attempting to measure 13 k with the 10 k range selected
The relative tolerance (%) display
of the measured value exceeds
the display range (999.999%).
Measuring 500 (+2400%) with a reference value of 20
The zero-adjusted value is out-
side of the display range.
Performing zero-adjustment after connecting 0.5 with the 1 range
Measuring 0.1 yields a -0.4 reading, exceeding the display
range.
While measuring, input voltage
exceed the A/D converter input
range.
Measuring a large resistance value in an electrically noisy environment
Current did not flow normally to
the measurement target.
(When the current fault mode set-
ting is set to “Over-range output”
only)
When the measurement target yields an open FAIL result
When either the SOURCE A or SOURCE B terminal suffers from poor
contact.
*To display “- - - - - - -” when a current fault occurs, set the current fault
mode setting to “Current fault.”(p.59)

Table of Contents

Other manuals for Hioki RM3545

Related product manuals