3.5 Checking Measured Values
58
LP OFF
LP ON
* When using the Z3003 Multiplexer Unit, ensure that the total of the unit’s internal wiring
resistance (including relays) and the wiring resistance from the connector to the measure-
ment target does not exceed the values in the above table.
You can verify that the unit’s internal wiring resistance is 1 or less using the self-test.
See: "8.6 Performing the Multiplexer Unit Self-test" (p.166)
Range
100 M range
high-precision
mode
Current
switching
Measurement
Current
SOURCE B - SOURCE A
(Other than measurement
target) *
10 m
−−1 A 1.5
100 m High 1 A 1.5
100 m
Low 100 mA 15
1000 m High 100 mA 15
1000 m
Low 10 mA 150
10 High 10 mA 150
10
Low 1 mA 1 k
100 High 10 mA 100
100
Low 1 mA 1 k
1000 1 mA 1 k
10 k
−−1 mA 1 k
100 k 100 A 1 k
1000 k
−−10 A1 k
10 M 1 A 1 k
100 M ON
100 nA 1 k
100 M OFF 1 A or less 1 k
1000 M OFF
1 A or less 1 k
Range
Measurement
Current
SOURCE B - SOURCE A
(Other than measurement target) *
1000 m 1 mA 2
10 500 A 5
100 50 A50
1000 5 A 500