10.4 External I/O Settings
206
The following external I/O settings are provided:
Input settings
Set the measurement start conditions (trigger source).(p.206)
Set the TRIG signal logic.(p.208)
Eliminate TRIG/PRINT signal chatter (filter function).(p.210)
Output settings
Set the EOM signal.(p.212)
Switch output modes (judgment mode/BCD mode).(p.214)
Measurements can be started in two ways.
10.4 External I/O Settings
Setting Measurement Start Conditions (Trigger Source)
Trigger signals are automatically generated internally for
continuous measurement.
Measure with internal (INT) triggering
To measure automatically
Measurements are triggered by an external signal.
Manual measurement triggering is also available.
Apply a trigger signal at the EXT I/O connector (p. 175)
Send the TRG command by communications interface
Press (only when EXT is selected)
Measure with external (EXT) triggering
To measure at specific times
To hold measured values as
desired
When internal triggering is enabled, the EXT I/O TRIG signal and the TRG command
are ignored (except for memory storage and statistical calculations).
To measure samples such as inductors that require time to settle, adjust delay time.
Start with a long delay, and gradually shorten it while watching for the measured value
to settle.
See: "4.9 Setting Pre-Measurement Delay" (p. 84)