8.4 Measuring with the Multiplexer
161
8
This section describes how to perform measurement while changing channels manually.
Before configuring these settings, see "Configuring Multiplexer Settings" (p.148) and "Cus-
tomizing Measurement Conditions for Individual Channels" (p.160).
8.4 Measuring with the Multiplexer
Measuring While Switching Channels Manually
1
Turn off the scan function.
See: "Configuring Multiplexer Settings" (p.148)
2
Change channels manually.
Measurement will be performed after applying the measurement conditions
for the selected channel. You can also change the measurement range,
speed, and comparator settings directly from the Measurement screen.
With the exception of channel operations, functionality is the same as measurement
using the terminals on the front of the instrument.
Select the channel.