6.6 Printing Measurement Data (as required)
114
Lower Value Lower limit value 50.0 A Lower allowable limit value*
Max Value Maximum value (259.9
A) –
Judgement Judgement result PASS Equal to or less than the allowable value
FAIL Greater than the allowable value
LOW Equal to or less than the Lower limit value.
--- PASS or FAIL is not determined.
Polarity Power supply
polarity
NORMAL Positive phase
REVERSE Negative phase
NONE None
Condition Equipment under
test status
NORMAL Normal condition
EARTH Single-fault condition
(disconnection of protective earth conductor)
POWERSOURCE Single-fault condition
(disconnection of one wire in power line)
NAPPLY Single-fault condition
(110% voltage application: positive phase)
RAPPLY Single-fault condition
(110% voltage application: negative phase)
LLINE Single-fault condition (line voltage application : L)
NLINE Single-fault condition (line voltage application : N)
Other
Condition
Other test condition
(Other applied
voltages)
NAPPLY
Single-fault condition (Positive phase applied)
RAPPLY
Single-fault condition (Negative phase applied)
NONE
Single-fault condition (No application)
Special
Condition
Special test condi-
tion (Specific
applied voltage)
NAPPLY
Single-fault condition (Positive phase applied)
RAPPLY
Single-fault condition (Negative phase applied)
NONE
Single-fault condition (No application)
S10 S10 status
ON S10 ON
OFF S10 OFF
S12 S12 status
ON S12 ON
OFF S12 OFF
S13 S13 status
ON S13 ON
OFF S13 OFF
Item Explanation Printed item Explanation