EasyManua.ls Logo

Hioki ST5540 - Page 182

Hioki ST5540
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.7 Command Message Description
176
bit7 Not used
Set all bits to 0.
Set to 0.
Set all bits to 0.
bit8 DC
Set at least one of bits 8 and 9 to
1.
bit9 AC
bit10
Application
of line
voltage L
Set all bits to 0.
bit11
Application
of line
voltage N
bit12
Not used
bit13
bit14
bit15
Bit Item
Explanation
1
*
2
*
3
*
4
*
5
*
*
1: Earth leakage current
*
2: Touch current: Enclosure - EarthTouch current: Enclosure - Enclosure, Free current
*
3: Patient auxiliary current, Patient leakage current: Patient - Earth,
Total patient leakage current: Patient - Earth
*
4: Patient leakage current: SIP/SOP, Total patient leakage current: SIP/SOP
*
5: Patient leakage current: F-type applied part, Patient leakage current: Metal accessible part
Total patient leakage current: F-type applied part,
Total patient leakage current: Metal accessible part