8.7 Command Message Description
177
8
Chapter 8 Controlling the Instrument From a Computer
Numerical value2
Bit Item
Explanation
1
*
2
*
3
*
4
*
5
*
bit0
Other test
condition
:Application of
110% volt-
age(none)
Set all bits to 0.
Set at least one
of bits 0, 1 and
2 to 1.
Set all bits to 0.
Set at least one
of bits 1 and 2
to 1.
Set all bits to 0.
bit1
Other test
condition
:Application of
110% volt-
age(Positive
phase)
bit2
Other test
condition
:Application of
110% volt-
age(Negative
phase)
bit3 Not used
Set all bits to 0. Set all bits to 0.
Set at least one
of bits 4 and 5 to
1.
bit4
Special test
condition
:Application of
110% volt-
age(Positive
phase)
bit5
Special test
condition
:Application of
110% volt-
age(Negative
phase)
bit6
Not used Set all bits to 0.
bit7
bit8
bit9
bit10
bit11
bit12
bit13
bit14
bit15
*
1: Earth leakage current
*
2: Touch current : Enclosure - Earth, Touch current: Enclosure - Enclosure
*
3: Patient auxiliary current, Patient leakage current: Patient - Earth,
Total patient leakage current: Patient - Earth
*
4: Patient leakage current: SIP/SOP, Total patient leakage current: SIP/SOP
*
5: Patient leakage current: F-type applied part, Patient leakage current: Metal accessible part
Total patient leakage current: F-type applied part,
Total patient leakage current: Metal accessible part