8.7 Command Message Description
198
When network B2 is selected
Example
Command
:CONFigure:AUTO OFF;:CONFigure:SPECial NAPPly
Specific manual measurement: Sets the application of 110% of rated voltage as
positive phase.
Query
Response
:CONFigure:AUTO OFF;:CONFigure:SPECial?
:CONFIGURE:SPECIAL NAPPLY (Headers : ON)
NAPPLY (Headers : OFF)
Specific manual measurement: The application of 110% of rated voltage has
been set to positive phase.
Error
An execution error occurs in the following cases:
When the measurement method is set to "automatic measurement"
When a mode other than "leakage current mode" is set
When a Network other than "B1" or "B2" is set
When a measurement mode other than Patient leakage current, F-type
applied part, Patient leakage current : Metal accessible part, Total patient
leakage current : F-type applied part, Total patient leakage current : Metal
accessible part is set
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.
Use this command after using the
:NETWork command to enter the "B1" or
"B2" network.
Use this command after executing the
:MODE command to activatePatient
leakage current:F-type applied part, Patient leakage current : Metal accessi-
ble part, Total patient leakage current : F-type applied part, Total patient leak-
age current : Metal accessible part.
Measurement modes
Special test condition (Specific applied voltage)
Class-I equipment Class-II equipment
Internally powered
equipment
Application
of 110%
voltage:
Positive
phase
Application
of 110%
voltage:
Negative
phase
Application
of 110%
voltage:
Positive
phase
Application
of 110%
voltage:
Negative
phase
Application
of 110%
voltage:Pos-
itive phase
Application
of 110%
voltage:
Negative
phase
Earth leakage current
Touch current :
Enclosure - Earth
––––––
Touch current :
Enclosure - Enclosure
––––––
Patient leakage
current :Patient - Earth
––––––
Patient leakage
current :SIP/SOP
––––––
Patient leakage
current :
F-type applied part

Patient leakage cur-
rent :
Metal accessible part

: Can be set, –: Cannot be set
Specific manual measurement: Sets and queries the application of 110% of
rated voltage(only ST5540)