8.7 Command Message Description
208
Error
An execution error occurs in the following cases:
• When network A, C, D, E, F or G is selected
• When a mode other than "leakage current mode" is set
• When network B1 or B2 is selected
When the measurement mode is selected (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).
Sets and queries the applied part of equipment under test
(Network B1, B2 only)(only ST5540)
Queries the event status register 0
Syntax
Query
Response
:ESR0?
<numerical value>
<numerical value> = 0 to 255 (NR1)
Function
Query Returns the setting content of event status register 0 (ESR0) as a numeric
value, then clears the content. No header is prefixed to the response message.
Example
Query
Response
:ESR0?
1
Bit 0 of ESR0 has been set to 1.
Error
If the response message exceeds the output queue size, a query error occurs.
1286432168421
bit 7bit 6bit 5bit 4bit 3bit 2bit 1bit 0
Not used ERROR LOW TEST MEAS T-FAIL FAIL PASS
Sets and queries the response header
Syntax
Command
Query
Response
:HEADer <characters>
:HEADer?
<characters>
<characters> = ON/OFF
ON : With response header
OFF : Without response header
Function
Command Sets the response header of the query. (default setting : OFF)
Query Returns the response header setting as alphabetic characters.
Example
Command
:HEADer OFF
Sets the response header to OFF.
Query
Response
:HEADer?
:HEADER ON (Headers : ON)
The response header has been set to ON.
OFF (Headers : OFF)
The response header has been set to OFF.