8.7 Command Message Description
209
8
Chapter 8 Controlling the Instrument From a Computer
Sets and queries the application line for leakage current measurement
between enclosure and line
Syntax
Command
Query
Response
:LINE <characters>
:LINE?
<characters>
<characters> = INT/ EXT
INT : Uses internal contact.
(internal contact and terminal T2)
EXT : Uses external contact.
(terminals T1 and T2)
Function
Command Sets the application line for Touch current Enclosure - Line or Enclosure - Line
leakage current.
Query Returns the setting of the application line for Touch current Enclosure - Line or
Enclosure - Line leakage current.
Example
Command
:LINE INT
Sets the application line for Touch current Enclosure - Line or Enclosure - Line
leakage current to "internal contact."
Query
Response
:LINE?
:LINE INT (Headers : ON)
INT (Headers : OFF)
Use of internal contact has been set.
Error
An execution error occurs in the following cases:
When the measurement mode is not set to Touch current Enclosure - Line,
Enclosure - Line leakage current.
When a mode other than "leakage current mode" is set
When the measurement method is set to "automatic measurement"(Com-
mand)
Clears the maximum value
Syntax
Command
:MAXimum:CLEar
Function
Command Clears the maximum value.
Example
Command
:MAXimum:CLEar
Clears the maximum value.
Error
An execution error occurs in the following cases:
When the measurement mode is not selected
When a mode other than "leakage current mode" is set
In automatic measurement
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.