8.7 Command Message Description
216
Error
If the set data unit number is greater than the total number of data units, an exe-
cution error occurs.
Queries the equipment name/control number of the saved data
Reads saved data
Syntax
Query
:MEMory:READ:MEASure? <numerical value>,<characters>
<numerical value> = Data unit number (NR1)
: Numeric value from 1 to the number of models (total number of
data units)
<characters> = Measurement mode
EARTh/ENCLosure1(ENCL1)/ ENCLosure2(ENCL2)/
ENCLosure3(ENCL3)/PATient1(PAT1)/
PATient2(PAT2)/PATient3(PAT3)/PAUXiliary/
TOUCh1(TOUC1)/TOUCh2(TOUC2)/
TOUCh3(TOUC3)/PATientP2E(PATP2E)/
PATientSIPSOP(PATSIPSOP)/
PATientFTYPE(PATFTYPE)/PATientMP(PATMP)/
TPATientP2E(TPATP2E)/TPATientSIPSOP(TPATSIPSOP)/TPA-
TientFTYPE(TPATFTYPE)/TPATientMP(TPATMP)/FREE
When network A, D, E, or F is selected
EARTh :Earth leakage current
ENCLosure1 :Enclosure - Earth leakage current
ENCLosure2 :Enclosure - Enclosure current
ENCLosure3 :Enclosure - Line leakage current
FREE :Free current (only network A)
When network B1 is selected(only ST5540)
EARTh :Earth leakage current
ENCLosure1 :Enclosure - Earth leakage current
ENCLosure2 :Enclosure - Enclosure current
PATient1 :Patient leakage current I
PATient2 :Patient leakage current II
PATient3 :Patient leakage current III
PAUXiliary :Patient auxiliary current
FREE :Free current
When network C or G is selected
EARTh :Earth leakage current
TOUCh1 :Touch current Enclosure - Earth
TOUCh2 :Touch current Enclosure - Enclosure
TOUCh3 :Touch current Enclosure - Line
When network B2 is selected(only ST5540)
EARTh :Earth leakage current
TOUCh1 :Touch current Enclosure - Earth
TOUCh2 :Touch current Enclosure - Enclosure
PATAUXiliary :Patient auxiliary current
PATientP2E :Patient leakage current Patient connection - Earth
PATientSIPSOP :Patient leakage current SIP/SOP
PATientFTYPE :Patient leakage current F-type applied part
PATientMP :Patient leakage current Metal accessible part
TPATientP2E :Total patient leakage current Applied part- Earth
TPATientSIPSOP :Total patient leakage current SIP/SOP
TPATientFTYPE :Total patient leakage current F-type applied part
TPATientMP :Total patient leakage current Metal accessible
part
FREE :Free current