1.6 Types of Leakage Current Measurement
17
1
Chapter 1 Overview
This is electric current that flows through an "applied part of malfunctioning medical equipment human
body
F-type applied part" path.
The measurement of this leakage current is required only for medical devices with an F-type applied part.
Instead of connecting damaged medical equipment to generate a single-fault condition, input a voltage
that is 110% of the rated voltage.
Patient leakage current measurement
(external voltage on a specific F-type applied part)