8.7 Command Message Description
224
Function
Command Sets the network.
Query Returns the network setting as alphabetic characters.
Example
Command
:NETWork A
Sets network A
Query
Response
:NETWork?
:NETWORK A (Headers : ON)
A (Headers: OFF)
Network A has been set.
Error
An execution error occurs in the following cases:
• When the measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).
Sets and queries the network
Start of automatic measurement
Syntax
Command
:STARt
Function
Command Starts measurement when automatic measurement is set.
Example
Command
:CONFigure:AUTO ON;:STARt
Starts automatic measurement.
Error
An execution error occurs in the following cases:
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• When polarity switching is set to power on setting and a power on check ends
in an error (NG), it is the instrument that is causing the error.
• When measurement time is set to 1 s.
• For Touch current Enclosure - Line and Enclosure - Line leakage current, if
the pre-check to prevent a ground fault results in NG judgement, an equip-
ment-dependent error occurs.
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.
Stops automatic measurements
Syntax
Command
:STOP
Function
Command Stops measurement when automatic measurement is set.
Example
Command
:STOP
Stops automatic measurement.
Error
An execution error occurs in the following cases :
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.