8.7 Command Message Description
223
8
Chapter 8 Controlling the Instrument From a Computer
Error
An execution error occurs in the following cases:
When a mode other than "leakage current mode" is set (Command)
When network is OFF(Command)
During automatic measurements(Command)
When power on polarity switching is set and a power on check ends in an
error (NG), it is the instrument that is causing the error.
When network A, C, D, E, or F is selected
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When set to Patient leakage current I, Patient leakage current II, Patient leak-
age current III, Patient auxiliary current, Patient leakage current Patient con-
nection - Earth, Patient leakage currentSIP/SOP, Patient leakage current F-
type applied part, Patient leakage current Metal accessible part, Total patient
leakage current Patient - Earth, Total patient leakage current SIP/SOP, Total
patient leakage current F-type applied part, Total patient leakage current
Metal accessible part mode
When an attempt is made to set enclosure - line leakage current mode and
touch current (Enclosure - Line) when the grounding class of equipment
under test is internally powered equipment
When network B1 is selected (only ST5540)
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When patient leakage current III mode is set for equipment under test whose
applied part is set to Type B
When patient leakage current II mode is set for a Type BF or Type CF applied
part
When
Enclosure - Line leakage current is set
When network B2 is selected(only ST5540)
When Earth leakage current measurement mode is set for equipment under
test whose grounding class is not Class I
When an attempt is made to set patient leakage current F-type applied part or
total patient leakage current F-type applied part for equipment under test
whose applied part is set to a B-type applied part
When an attempt is made to set patient leakage current metal accessible part
or total patient leakage current metal accessible part for a CF-type applied
part
Sets and queries the measurement mode
Sets and queries the network
Syntax
Command
Query
Response
:NETWork <characters>
:NETWork?
<characters>
<characters> = A/ B2/ C/ D/ E/ F/ G/B1/OFF
A: Network A
B2: Network B2(only ST5540)
C: Network C
D: Network D
E: Network E
F: Network F
G: Network G
B1: Network B1(only ST5540)
OFF: No network setting