EasyManua.ls Logo

Hioki ST5540 - Page 232

Hioki ST5540
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.7 Command Message Description
226
Sets and queries the beep sound (Key entry)
Syntax
Command
Query
Response
:SYSTem:BEEPer:KEY <characters>
:SYSTem:BEEPer:KEY?
<characters>
<characters> = ON/OFF
ON : Generates a beep sound.
OFF:Does not generate a beep sound.
Function
Command Sets the beep sound for notification of key entry.
Query Returns the beep sound setting for notification of key entry.
Example
Command
:SYSTem:BEEPer:KEY ON
Sets the beep sound to ON.
Query
Response
:SYSTem:BEEPer:KEY?
:SYSTEM:BEEPER:KEY ON (Headers : ON)
ON (Headers : OFF)
The beep sound has been set to ON.
Error
An execution error occurs in the following cases:
When the measurement mode is selected (Command)
When a mode other than "leakage current mode" is set (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).
Sets and queries the beep sound (T2 output)
Syntax
Command
Query
Response
:SYSTem:BEEPer:T2OUt <characters>
:SYSTem:BEEPer:T2OUt?
<characters>
<characters> = ON/OFF
ON : Generates a beep sound.
OFF:Does not generate a beep sound.
Function
Command Sets the beep sound to be generated while line voltage is output from terminal
2.
Query Returns the beep sound setting for notification of line voltage output from termi-
nal T2.
Example
Command
:SYSTem:BEEPer:T2OUt ON
Sets the beep sound to ON.
Query
Response
:SYSTem:BEEPer:T2OUt?
:SYSTEM:BEEPER:T2OUT ON (Headers : ON)
ON (Headers : OFF)
The beep sound has been set to ON.
Error
An execution error occurs in the following cases:
When the measurement mode is selected (Command)
When a mode other than "leakage current mode" is set (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).

Table of Contents