8.7 Command Message Description
227
8
Chapter 8 Controlling the Instrument From a Computer
Sets and queries the beep sound (T3 output)
Syntax
Command
Query
Response
:SYSTem:BEEPer:T3OUt <characters>
:SYSTem:BEEPer:T3OUt?
<characters>
<characters> = ON/OFF
ON: Generates a beep sound.
OFF :Does not generate a beep sound.
Function
Command Sets the beep sound to be generated while 110% voltage is output from termi-
nal 3.
Query Returns the beep sound setting for notification of 110% voltage output from ter-
minal T3.
Example
Command
:SYSTem:BEEPer:T3OUt ON
Sets the beep sound to ON.
Query
Response
:SYSTem:BEEPer:T3OUt?
:SYSTEM:BEEPER:T3OUT ON (Headers : ON)
ON (Headers : OFF)
The beep sound has been set to ON.
Error
An execution error occurs in the following cases:
• When the measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).
Clears the panel
Syntax
Command
:SYSTem:CLEar
Function
Command Initializes all test condition setting data including the contents of all panels.
Example
Command
:SYSTem:CLEar
Initializes all panels.
Error
An execution error occurs in the following cases :
• When the measurement mode is selected
• When a mode other than "leakage current mode" is set
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).
Sets and queries the date
Syntax
Command
Query
Response
:SYSTem:DATE <numerical value 1>,<numerical value 2>,<numerical
value 3>
:SYSTem:DATE?
<numerical value 1>,<numerical value 2>,<numerical value 3>
<numerical value 1> = Year : 2000 to 2099 (NR1)
<numerical value 2> =
Month : 1 to 12 (NR1)
<numerical value 3> = Day : 1 to 31 (NR1)