8.7 Command Message Description
235
8
Chapter 8 Controlling the Instrument From a Computer
Sets and queries the display language
Syntax
Command
Query
Response
:SYSTem:LANGuage <characters>
:SYSTem:LANGuage?
<characters>
<characters> = JAPanese/ ENGlish
JAPanese : Japanese language
ENGlish : English language
Function
Command Sets the display language.
Query Returns the display language setting.
Example
Command
:SYSTem:LANGuage JAPanese
Sets the display language to Japanese.
Query
Response
:SYSTem:LANGuage?
:SYSTEM:LANGUAGE JAPANESE (Headers : ON)
JAPANESE (Headers : OFF)
The display language has been set to Japanese.
Error
An execution error occurs in the following cases :
When the measurement mode is selected (Command)
When a mode other than "leakage current mode" is set (Command)
Note
Use this command after executing the :MODE command to activate OFF
mode (i.e., measurement mode not selected).
Loads a panel
Syntax
Command
:SYSTem:LOAD <numerical value>
<numerical value> = 1 to 30 (NR1)
Function
Command Loads (reads) the content of the panel of the specified number.
Example
Command
:SYSTem:LOAD 10
Loads the content of panel number 10.
Error
An execution error occurs in the following cases:
When the measurement mode is selected
If the designated panel number does not conform to the specified range
When the panel of the specified number is not stored
When a mode other than "leakage current mode" is set
Note
Use this command after executing the :MODE command to activate OFF
mode (i.e., measurement mode not selected).
When power on polarity switching is set and a power on check ends in an
error (NG), it is the instrument that is causing the error.
Sets and queries all allowable values (lower limit) On/Off
Syntax
Command
Query
Response
:SYSTem:LOWer <characters>
:SYSTem:LOWer?
<characters>
<characters>=OFF/ON
OFF: All allowable values (lower limit) OFF
ON: All allowable value (lower limit) ON