8.7 Command Message Description
238
Initializes the ST5540/ST5541
Syntax
Command
:SYSTem:RESet <characters>
<characters> = ALL/ CONDition/ SAVEdata
ALL :Initializes the instrument to the default
settings (factory settings).
(Same as *RST command)
CONDition :Initializes the instrument to the default settings except for
saved measurement data.
SAVEdata :Initializes only the saved measurement data.
Function
Command Initializes the HiTESTER.
After initialization, the initial screen is displayed.
Example
Command
:SYSTem:RESet SAVEdata
Initializes only the saved measurement data.
Note
The communication conditions are not initialized.
Saves a panel
Syntax
Command
:SYSTem:SAVE <numerical value>
<numerical value> = 1 to 30 (NR1)
Function
Command Saves (stores) the present settings in the panel of the specified number.
Example
Command
:SYSTem:SAVE 5
Saves setting data in panel number 5.
Error
An execution error occurs in the following cases :
• When the measurement mode is not selected
• If the designated panel number does not conform to the specified range
• When a mode other than "leakage current mode" is set
Queries the fuse check
Syntax
Query
Response
:SYSTem:TEST:FUSE?
<characters>
<characters>=OK/NG
OK: FUSE CHECK OK
NG: FUSE CHECK NG
Function
Query Performs a fuse check and returns the result as alphabetic characters.
Example
Error
Query
Response
:SYSTem:TEST:FUSE?
:SYSTEM:TEST:FUSE OK (Headers : ON)
OK (Headers : OFF)
The result of the fuse check is OK.
An execution error occurs in the following cases :
• When measurement mode is selected
• When a mode other than "leakage current mode" is set
Note
• Use this command after using the :MODE command to turn off
(and no measurement mode is selected).