8.7 Command Message Description
241
8
Chapter 8 Controlling the Instrument From a Computer
Sets and queries the VA check (At power on)
Syntax
Command
Query
Response
:SYSTem:TEST:VA:STARt <characters>
:SYSTem:TEST:VA:STARt?
<characters>
<characters>=ON/OFF
ON : Performs VA check at power on.
OFF : Does not perform VA check at power on.
Function
Command
Sets VA check at power on.
Example
Command :SYSTem:TEST:VA:STARt ON
Performs VA check at power on.
Query
Response
:SYSTem:TEST:VA:STARt?
:SYSTEM:TEST:VA:START ON (Headers: ON)
ON (Headers: OFF)
Performs VA check at power on.
Sets and queries the time
Syntax
Command
Query
Response
:SYSTem:TIME <numerical value 1>,<numerical value 2>
:SYSTem:TIME?
<numerical value 1>,<numerical value 2>
<numerical value 1> = Time 0 to 23 (NR1)
<numerical value 2> = Minute 0 to 59 (NR1)
Function
Command Sets the time in order of hours and minutes.
Query Returns the time setting in numeric values.
Example
Command
:SYSTem:TIME 12,34
Sets the time to 12:34.
Query
Response
:SYSTem:TIME?
:SYSTEM:TIME 12,34 (Headers : ON)
12,34 (Headers : OFF)
The time has been set to 12:34.
Error
An execution error occurs in the following cases :
When the measurement mode is selected (Command)
If the set numeric value is not within the specified range (Command)
When a mode other than "leakage current mode" is set (Command)
If the character string does not conform to the specified format (Command)
Note
Use this command after executing the :MODE command to activate OFF mode
(i.e., measurement mode not selected).