Appendix3 List of default settings
A13
Appendix
Network B2
After selecting network, class, applied part and leakage current mode
When network B2 is selected, measured AC+DC current does not appear on the
screen. AC+DC is automatically selected internally.
default setting
Class I Class II Internally powered
Leakage
current mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
Earth leak-
age current
measurement manual
––––––
polarity positive phase
equipment sta-
tus
normal
Other applied
voltages
–
Application of
specific voltage
–
s10 OFF
s12 OFF
s13 –
filter ON
target current ACDC
range AUTO
allowable value
: upper limit :
normal
5.000 mA
allowable value
: upper limit :
fault
10.00 mA
allowable value
: lower limit :
ON/OFF :
normal
OFF
allowable value
: lower limit :
ON/OFF : fault
OFF
*
( ) is the value after DC switching