Appendix3 List of default settings
A21
Appendix
Total patient
leakage
current
( F-type ap-
plied part)
measurement
manual
manual
manual
polarity positive phase positive phase
equipment
status
–––
Other applied
voltages
–––
Application of
specific voltage
110%
applied N
110%
applied N
110%
applied N
s10 OFF OFF
s12
s13 OFF OFF
filter ON ON ON
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value
: upper limit:
normal
–––
allowable value:
upper limit: fault
5.000
mA
100.0
A
5.000
mA
100.0
A
5.000
mA
100.0
A
allowable value:
lower limit :
ON/OFF:
normal
–––
allowable value:
lower limit: ON/
OFF: fault
OFF OFF OFF
Total patient
leakage
current
(metal ac-
cessible
part)
measurement manual
manual
manual
polarity positive phase positive phase
equipment
status
–––
Other applied
voltages
–––
Application of
specific voltage
110%
applied N
110%
applied N
110%
applied N
s10 OFF OFF
s12
s13
filter ON ON ON
target current ACDC ACDC ACDC
rangeAUTOAUTOAUTO
allowable value
: upper limit :
normal
–––
allowable value
: upper limit :
fault
1.000 mA 1.000 mA 1.000 mA
allowable value
: lower limit :
ON/OFF :
normal
–––
allowable value
: lower limit :
ON/OFF : fault
OFF OFF OFF
default setting
Class I Class II Internally powered
Leakage
current mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
*
( ) is the value after DC switching