Appendix3 List of default settings
A22
Patient auxil-
iary current
measurement manual manual manual
polarity positive phase positive phase
equipment
status
normal normal normal
Other applied
voltages
–––
Application of
specific voltage
–––
s10 OFF OFF
s12
s13
filter ON ON ON
target current AC AC AC
range AUTO AUTO AUTO
allowable value
: upper limit :
normal
*
100.0 A
(10.00 A)
10.00
A
(10.00
A)
100.0 A
(10.00 A)
10.00
A
(10.00
A)
100.0 A
(10.00 A)
10.00
A
(10.00
A)
allowable value
: upper limit :
fault
*
500.0 A
(50.00
A)
50.00
A
(50.00
A)
500.0
A
(50.00
A)
50.00
A
(50.00
A)
allowable value
: lower limit :
ON/OFF :
normal
OFF OFF OFF
allowable value
: lower limit :
ON/OFF : fault
OFF OFF
default setting
Class I Class II Internally powered
Leakage
current mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
*
( ) is the value after DC switching