3.5 Pre-Test Inspection
46
(4) Polarity check results
The wiring may not be configured for uninterrupted polarity
switching. Check the wiring.
The grounding wire may be disconnected.
The power supply may be a floating power supply.
The power supply may be a single-phase, 3-wire power
source.
The polarities of the power supply are reversed.
Reverse the L and N wire connections on [LINE IN].
See "Connecting the power cord of the equipment under test" (p. 33)