4.4 Setting the Measurement Range (Auto/Hold)
56
Range of indication depends on selected unit of indication.
See "7.1 Setting the Mode (To use the instrument as a voltmeter)" (p. 117)
See For details of accuracy guaranteed range: "10.4 Accuracy" (p. 258)
See "7.10 Setting Indication Unit" (p. 131)
When selecting network A, B1, B2, C, E, or G
(Measurement of AC, DC, or AC+DC)
(Measurement of ACPeak)
When selecting network D
(Measurement of AC, DC, or AC+DC)
(Measurement of ACPeak)
When selecting network F
(Measurement of AC, DC, or AC+DC)
(Measurement of ACPeak)
Range 50.00 mA 5.000 mA 500.0 A 50.00 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
50.00 mA
|
0.00 mA
(0.00 mA)
5.000 mA
|
0.000 mA
(0.000 mA)
500.0 A
|
0.0 A
(0.0 A)
50.00 A
|
1.01 A
(0.00 A)
Range 75.0 mA 10.00 mA 1.000 mA 500.0 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
75.0 mA
|
0.0 mA
(0.0 mA
or less)
10.00 mA
|
0.00 mA
(0.00 mA or less)
1.000 mA
|
0.000 mA
(0.000 mA or less)
500.0
A
|
1.1
A
(0.0
A)
Range 50.00 mA 5.000 mA 500.0 A 50.00 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
33.00 mA
|
0.00 mA
(0.00 mA)
3.300 mA
|
0.000 mA
(0.000 mA)
330.0 A
|
0.0 A
(0.0 A)
33.00 A
|
1.01 A
(0.00 A)
Range 75.0 mA 10.00 mA 1.000 mA 500.0 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
50.00 mA
|
0.0 mA
(0.0 mA or less)
6.60 mA
|
0.00 mA
(0.00 mA or less)
0.660 mA
|
0.000 mA
(0.000 mA or less)
330.0 A
|
1.1 A
(0.0 A)
Range 50.00 mA 5.000 mA 500.0 A 50.00 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
25.00 mA
|
0.00 mA
(0.00 mA)
2.500 mA
|
0.000 mA
(0.000 mA)
250.0 A
|
0.0 A
(0.0 A)
25.00 A
|
1.01 A
(0.0 A)
Range 75.0 mA 10.00 mA 1.000 mA 500.0 A
Maximum indication value
to
Minimum indication value
(In Hold range setting)
37.5 mA
|
0.0 mA
(0.0 mA or less)
5.00 mA
|
0.00 mA
(0.00 mA or less)
0.500 mA
|
0.000 mA
(0.000 mA or less)
250.0 A
|
1.1 A
(0.0 A)