Measuring Components or Circuits
When automatically measuring an object with a large capacitance (such as a
capacitor)
Improving the reliability of insulation resistance measurement
Set the contact check function that checks connections (contacts) in the measurement circuit
including the measured object and the measurement xture.
See “4.12 Contact Check (Executing Contact Check, Setting Reference Value)” (p. 71).
Reducing the measurement time and improving the measurement accuracy
Pre-charge the measured object before measurement.
Connection to the CHARGE terminal (p. 33) and the following settings are required:
See “4.8 Setting Pre-charge” (p. 64) and “4.9 Limiting Current Flowing Through Measured Object
(Current Limiter)” (p. 65)
INPUT
GUARD
GROUND
CHARGE
OUTPUT
A
SM7110/SM7120
Measured
object
Charging
Measured
object
Negative side Positive side
Rx
Rc
Shorting plug
Measurement
voltage
source
+
Ammeter
Current limiter
Negative
side
Positive
side
Rc is the measured object after Rx.
Example of CHARGE terminal connection (A grounded measured object)
5
Measurement Methods Suitable for Various Objects to Be Measured