5-16 
5.10  Mist collector maintenance(applied UX-E only)   
1  Remove head cover and heater cover. 
 
2  Unscrew the fixing screw of coupling cover, and then remove the coupling cover. 
 
 
 
 
 
 
 
 
 
 
3  Remove the nut of the mist collector, and then using tweezers, take out O-ring and 
the filter from inside. After that, attach a new O-ring P-6EPT and a new PTFE filter 
F-20-8.8. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
4  Attach the removed nut to mist collector, and then attach all removed parts in the   
reverse order of the procedures above. 
 
Fixing screw of coupling cover (1 location) 
Nut 
*Using two 8 mm wrenches, hold   
housing by one wrench and remove the nut 
by the other wrench.