6-44
6.4.3 Viscosity Readings Out of Range
Viscosity Reading
Out of Range
Resume the
operation.
Is it normally
measured?
Yes
No
Perform the viscometer
calibration
Perform the viscometer
operation test on the
Circulation system control
screen.
Disconnect and reconnect the
following connectors.
EZJ126: CN13, CN22
EZJ132: CN1 to 4
Proximity sensor, MV10
Disassemble and clean the
viscometer.
Replace the viscometer
Is it normally measured?
Is it normally measured?
Yes
Yes
No
No
Replace the ink.
Is it normally measured?
Yes
No