5A5A
OPERATING INSTRUCTIONS
5A.22
ThinPrep 2000 Processor Operator’s Manual
1 – Counters:
Pressing
1
displays the Sequence Counters. A number is displayed next to each sequence number
which identifies the number of cycles for that particular sequence. The value next to the “T” is the
total number of cycles on the processor. To return to the Status Menu, press STOP.
2 – Error History:
Pressing
2
displays the Error History screen. The system will store the last 50 error messages that
occurred on the processor. Technical Service may ask you to access this screen during troubleshoot-
ing. The first column (#) is the counter, 1–50. The second column (ERROR) is the error code. The third
column (MINOR) is the minor error number which often provides additional information on the
source of the error. The last column (CYCLE) is the total cycle count of the processor when the error
occurred. To return to the Status Menu, press STOP.
Sequence Counters:
1 - XXXXXX 4 - XXXXXX
2 - XXXXXX
3 - XXXXXX T - XXXXXX
Error History: ↑↓
# ERROR MINOR CYCLE
XX XX XX XXXXXX