NON-GYNECOLOGIC SAMPLE PREPARATION
1.30
ThinPrep 2000 Processor Operator’s Manual
Mucoid Specimens
Problem Procedure
A. Did the “Sample Is
Dilute” message
appear during pro-
cessing?
1. Check to see if cellularity is adequate. If
not, use more of the pellet if available.
Prepare a slide using Sequence 3.
B. Does the slide have an
obvious “halo” of cel-
lular material and/or
white blood cells?
1. Dilute the sample 20:1. Use a calibrated
pipette to add 1 mL of sample to a new
PreservCyt Solution vial. Prepare slide
using Sequence 1.
If a halo is present on the new slide, call
Hologic Technical Service.
C. Is the slide sparse and
does it contain mucus?
1. Pour the contents of the PreservCyt sam-
ple vial into a centrifuge tube.
Call Hologic Technical
Service.
2. Concentrate by centrifugation.
— 600g x 10 min.
3. Pour off supernatant and vortex to
resuspend cell pellet.
4. CytoLyt Solution Wash
5. Evaluate cell pellet appearance.
Refer to page 1.11.
If pellet contains mucus, repeat from
step 4.